Stryker and FDA notified healthcare professionals of a recall of 23 Operating Room System II Surgical Navigation Systems because there is a potential for the navigation PC SPC-1 component to stop working which could result in the screen freezing, the system updating at a slow rate, or not responding at all. The Navigation System II is a computer aided surgery platform that surgeons can use to perform Hip, Knee, Spine, Neuro and ENT surgical procedures and contains a computer workstation with the navigation System II software and various components necessary to run the system.The potential harms associated with this failure are: delay in surgery, reschedule of the procedure resulting in an additional surgery, risk of infection, increased morbidity, potential neurological deficits, or injury due to the surgeon operating in an area where they did not intend to operate. Depending on the type of surgery, these failures could potentially lead to serious adverse health consequences, including death. Hospitals that have product that corresponds to the catalog numbers above should immediately quarantine the product, label it as a recalled product and stop using the product.
Archive for: 2009
Cardiovascular Systems, Inc. and FDA notified healthcare professionals of a nationwide recall of all lots of the ViperSheath Sheath Introducer on behalf of Thomas Medical due to reports about stretching or fracture of the sheath during use.
In the event of a device fracture, separated segments of the device may require unplanned open surgery to remove the retained segments or control bleeding. Since this device is coil reinforced, any separation of the cannula (a flexible tube inserted into the body) has the potential to expose portions of the coil, creating the potential for vessel dissection or perforation.
The recall encompasses products distributed from March 25, 2009 to October 21, 2009, and includes the following lot ranges and catalog numbers:
- Lot range: S28117 through S29174
- Catalog numbers: VPR-ISH 5 X 85, VPR-ISH 6 X 85, VPR-ISH 7 X 85, VPR-ISH 5 X 45, VPR-ISH 6 X 45, VPR-ISH 7 X 45
FDA notified healthcare professionals of a Class I Recall of all lots of the Synthes USA, Ti Synex II Vertebral Body Replacement, a device used in the T1-L5 portion of the spine to replace a collapsed, damaged, or unstable vertebral body. Reports of moderate to severe loss of vertebral body replacement height (caused by failure of the central body component) in situ at six to fifteen months post implantation were received. Potential adverse health issues that could be associated with this issue include neural injury, increased pain, spinal kyphosis if unrecognized, failure of supplementary fixation, and/or need for reoperation/revision surgery. Surgeons and hospitals in possession of the subject devices must stop implanting them immediately. This product was manufactured from June 8, 2007 through September 9, 2009 and distributed from July 2, 2007 through September 8, 2009.
FDA and Edwards Lifesciences notified healthcare professionals about the Class 1 recall of CardioVations EndoClamp Aortic Catheter, Model Numbers EC1001 and EC65, a device that blocks off the aorta, monitors aortic pressure, and delivers solution to stop the heart during cardiopulmonary bypass procedures. The recall was initiated because the balloon catheters may spontaneously rupture during surgery. This product was manufactured from August, 2008 through August, 2009 and distributed from November, 2008 through September, 2009.
Centurion Medical Products and FDA notified healthcare professionals about a Class I recall of Premie Pack, Kit Code LM 110 and Full Term Meconium Pack, Kit Code LM115. The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated on the label, which could result in an inability to remove secretions and cause partial or complete blockage of the airway and the inability to ventilate the patient. These products were distributed from January 1, 2007 through September 22, 2009.
Pointe Scientific and FDA notified healthcare professionals of a nationwide recall of all size kits of Liquid Glucose Hexokinase Reagent catalog number G7517. The reagents have been found to fail linearity at >200mg/dL that results in inaccurate glucose values above this range. Distributors and testing laboratories who have received the Liquid Glucose Hexokinase Reagent (G7517) which is being recalled should destroy remaining inventory.
Testing laboratories should consider all test results obtained with the lot numbers listed above to be questionable. The laboratory should inform the patient’s attending physician, and determine, with their input, whether confirmation of the previous test results will be required.
Qualitest Pharmaceuticals Issues a Nationwide Recall of All Accusure® Insulin Syringes
Qualitest Pharmaceuticals today issued a voluntary nationwide recall of all Accusure® Insulin Syringes. The distributed syringes are of the following descriptions and NDC numbers: 28G 1/2cc, NDC 0603-6995-21;28G 1cc, NDC 0603-6996-21; 29G 1/2cc NDC 0603-6997-21, 29G 1cc, NDC 0603-6998-21, 30G 1/2cc, NDC 0603-999-21, 30G 1cc, NDC 0603-7000-21, 31G 1/2cc, NDC 0603-7001-21; and 31G 1cc, NDC 0603-7002-21. All Accusure® Insulin Syringes regardless of lot number are subject to this recall. These syringes were distributed between January 2002 and October 2009 to wholesale and retail pharmacies nationwide (including Puerto Rico). The syringes in these lots may have needles which detach from the syringe.
If the needle becomes detached from the syringe during use, it can become stuck in the insulin vial, push back into to the syringe, or remain in the skin after injection.
Consumers who have any Accusure® Insulin Syringes should stop using them and contact Qualitest at 1-800-444-4011 for reimbursement. You can find the lot number on the white paper backing of each individual syringe.
Qualitest is notifying all customers who received these syringes and arranging for the return of any affected product.
The recall is being made with the knowledge of the Food and Drug Administration.
Consumers with questions may contact Qualitest at 1-800-444-4011 for more information.
Adverse reactions or quality problems experienced with the use of this product may be reported to FDA’s MedWatch Adverse Event Reporting program either on line, by regular mail or by fax.
- Online: www.fda.gov/medwatch/report.htm
- Regular Mail: use postage-paid FDA form 3500 available at: www.fda.gov/medwatch/getforms.htm
Mail to MedWatch 5600 Fishers Lane, Rockville, MD 20852-9787 - Fax: 1-800-FDA-0178
American Regent Voluntarily Recalls All Lots of Ketorolac Tromethamine Injection, USP 30 mg/mL; 1mL and 2mL Single Dose Vials
American Regent conducts nationwide voluntary recall of ALL lots of its Ketorolac Tromethamine Injection, USP 30 mg/mL:
- NDC# 0517-0801-25 30 mg/mL 1mL Single Dose Vial
- NDC# 0517-0902-25 30mg/mL 2mL Single Dose Vial (60mg/2mL)
This voluntary recall is due to the potential that particulate matter in conjunction with crystallization may be present in the product. This recall does not include other concentrations of AMERICAN REGENT Ketorolac Tromethamine Injection, USP. American Regent is undertaking this recall in consideration of the potential for safety issues if the product is administered to patients, including obstruction of blood vessels which can induce pulmonary emboli or thrombosis, activate platelets and/or neutrophils to induce anaphylactic reactions. Other adverse effects associated with the injection of particulate matter include foreign body granulomas, and local irritation at the injection site.
The product was distributed to wholesalers and distributors nationwide and to one account in Abu Dhabi.
Hospitals, surgi-centers, clinics and other healthcare facilities should not use any AMERICAN REGENT Ketorolac Tromethamine Injection, USP Injection 30 mg/mL for patient care and should immediately quarantine any product for return.
“Patient safety is our primary concern, and we are committed to taking the necessary steps to protect patients from any potential safety risks,” said Mary Jane Helenek, President and CEO of American Regent.
On Friday, October 16, 2009, American Regent voluntarily recalled ALL unexpired lots of Ketorolac Tromethamine Injection, USP, 30 mg/mL due to the presence of particulate matter in conjunction with crystallization. This recall does not affect the other strength of Ketorolac Tromethamine Injection, USP, 15mg/mL, 1 mL Single Dose Vial, NDC # 0517-0601-25.
While American Regent continues to investigate this issue, the company is taking precautionary action and initiated this voluntary recall. American Regent has informed the FDA of its actions and is maintaining ongoing discussions with the Agency.
As is standard practice, and as stated in the Ketorolac Tromethamine Injection Product Package Insert, “Parenteral drug products should be inspected visually for particulate matter and discoloration prior to administration whenever solution and container permit.”
American Regent will credit accounts for all returned Ketorolac Tromethamine Injection, USP product. Those with questions about the return process, please call our Customer Service Department at 1-800-645-1706: Monday thru Friday from 8:30 AM to 7:00 PM ET.
Hospitals, surgi-centers, clinics and healthcare providers, or patients with other questions may contact the Professional Services Department at 631-924-4000.
Any adverse reactions experienced with the use of this product, and/or quality problems should be reported to American Regent, Inc. via email at PV@luitpold.com, by fax to 610-650-7781 or 610-650-0170 or by phone at 1-800-734-9236. Adverse reactions may also be reported to FDA’s MedWatch Adverse Event Reporting program online, by phone [1-800-FDA-1088], or by returning the postage paid FDA form 3500, by mail to [MedWatch, 5600 Fishers Lane, Rockville, MD 20852-9787] or via fax [1-800-FDA-0178].
Ketorolac Tromethamine Injection, USP is manufactured by Luitpold Pharmaceuticals, Inc. and is distributed by American Regent, Inc. (Shirley, NY).
Unomedical Inc., a manufacturer of medical devices, today announced that it is conducting a voluntary recall of certain units of the single-patient use Manual Pulmonary Resuscitator (MPR). The recall only impacts MPRs manufactured from July 2002 – March 2008 and matching the lot numbers listed on the following Unomedical web page: Unomedical MPR Recall. This recall is being conducted because of a potential malfunction of certain units of the MPR, which may impair the ability of the device to generate the positive pressure necessary to function properly. The occurrence of such a malfunction may create a situation in which the use of the product could potentially cause serious adverse health consequences or death. This recall does not impact any MPRs manufactured after March 2008.
Unomedical is contacting customers to arrange for the return and credit of all MPR units subject to this recall by sending notification letters to distributors and customers. In addition, the company has set up a web page with a list of affected lot numbers, guidance (diagram and photos) to allow customers to identify products subject to the recall in the event that the customer has already removed the primary product packaging, and instructions on what actions to take.
In order to distinguish between the recalled product and unaffected product, customers may examine the patient valve housing immediately below the right-angle exhalation port, where the retention ring should be visible. MPR units with a clear or transparent ring, as well as those where no ring can be seen, should be returned to Unomedical as instructed. MPR units with a clearly visible blue retention ring are not affected and do not need to be returned.
The MPR is a single-patient use device used by healthcare professionals. It is intended for patients requiring total or intermittent ventilatory support.
Unomedical notified the U.S. Food & Drug Administration (FDA) of its decision to voluntarily recall the product on October 1, 2009.
Customers with questions are urged to contact Unomedical at 1-800-634-6003. Any adverse reactions experienced with the use of this product, and/or quality problems can also be reported to the FDA’s MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF- 2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch website at: www.fda.gov/medwatch.
New USP Standards for Heparin Products Will Result in Decreased Potency
Adjustments may be Needed to Achieve Desired Anticoagulant Effect in Some Patients
New Heparin to Ship Starting October 8
The U.S. Food and Drug Administration today alerted health care professionals to a change in heparin manufacturing that is expected to decrease the potency of the common anti-clotting drug.
To ensure the quality of heparin and to guard against potential contamination, the United States Pharmacopeia (USP), a nonprofit standards-setting organization, adopted new manufacturing controls for heparin. These changes include a modification of the reference standard for the drug’s unit dose.
Manufacturers in the United States label the amount of heparin included in their products based on USP standards. The changes adopted by the USP for the heparin unit dose match the World Health Organization’s International Standard (IS) unit dose definition that has been in use in Europe for many years. The revised USP reference standard and unit definition for heparin is about 10 percent less potent than the former USP unit.
A unit is the measure of a drug’s activity in the body. For heparin, a unit dose is the measure of the drug’s ability to block the blood’s natural clotting ability (anticoagulation). Heparin’s potency is determined by the dose of the drug required to produce a specific level of anticoagulation.
Manufacturers for the U.S. market have begun to make heparin using the new USP standard. While the USP manufacturing controls take effect Oct. 1 for production, the FDA has asked that they not ship this new product to customers until Oct. 8, 2009, or later. The delay will give health care providers and facilities time to learn about the changes and to make adjustments to their pharmacy procedures and dosing practices, according to John Jenkins, M.D. director of the Office of New Drugs in the FDA’s Center for Drug Evaluation and Research.
“Although the FDA-approved labeling for heparin has not changed, including the recommended doses, it is essential that health care professionals be aware of the potential difference in potency between the old and new vials of heparin when administering the drug,” said Jenkins.
Four companies market heparin in the United States. APP, the largest manufacturer, markets heparin in vials; Hospira markets heparin in intravenous bags, vials, and syringes; Baxter markets heparin in intravenous bags, and B. Braun markets heparin in intravenous bags. The FDA has asked that all manufacturers identify their new products to help pharmacies and health care professionals differentiate it from the former product.
Prescription and over–the–counter medicines available in the United States must generally meet USP’s public standards, when such standards exist. The revised standards for heparin are contained in a new USP monograph.
The monograph was revised, in part, in response to a 2007- 2008 incident of heparin contamination involving a manufacturing step in China. The contaminated heparin was associated with deaths and other adverse events in the United States. The monograph was changed to include a test for the contaminant.