Stryker and FDA notified healthcare professionals of a recall of 23 Operating Room System II Surgical Navigation Systems because there is a potential for the navigation PC SPC-1 component to stop working which could result in the screen freezing, the system updating at a slow rate, or not responding at all. The Navigation System II is a computer aided surgery platform that surgeons can use to perform Hip, Knee, Spine, Neuro and ENT surgical procedures and contains a computer workstation with the navigation System II software and various components necessary to run the system.The potential harms associated with this failure are: delay in surgery, reschedule of the procedure resulting in an additional surgery, risk of infection, increased morbidity, potential neurological deficits, or injury due to the surgeon operating in an area where they did not intend to operate. Depending on the type of surgery, these failures could potentially lead to serious adverse health consequences, including death. Hospitals that have product that corresponds to the catalog numbers above should immediately quarantine the product, label it as a recalled product and stop using the product.
Archive for: December, 2009
Cardiovascular Systems, Inc. and FDA notified healthcare professionals of a nationwide recall of all lots of the ViperSheath Sheath Introducer on behalf of Thomas Medical due to reports about stretching or fracture of the sheath during use.
In the event of a device fracture, separated segments of the device may require unplanned open surgery to remove the retained segments or control bleeding. Since this device is coil reinforced, any separation of the cannula (a flexible tube inserted into the body) has the potential to expose portions of the coil, creating the potential for vessel dissection or perforation.
The recall encompasses products distributed from March 25, 2009 to October 21, 2009, and includes the following lot ranges and catalog numbers:
- Lot range: S28117 through S29174
- Catalog numbers: VPR-ISH 5 X 85, VPR-ISH 6 X 85, VPR-ISH 7 X 85, VPR-ISH 5 X 45, VPR-ISH 6 X 45, VPR-ISH 7 X 45